Pathological staining is one of the most commonly used diagnostic methods in clinical diagnosis and research experiments. Among the frequently used techniques, paraffin sectioning and staining with liquid are commonly employed.

Highlights

Only by ensuring the quality of each step in the production process can we obtain the desired pathological staining slides, thus obtaining precise diagnostic results and further experimental data. However, in experiments, the quality of pathological slides tends to vary due to various factors. Therefore, after collecting various relevant literature, based on experimental experience, some common problems that may arise in paraffin sectioning and liquid staining are summarized, then to discuss the reasons in-depth for each occurrence and seek targeted solutions.

Since there are various reasons for errors in experimental results, the causes and solutions provided for the common problems with paraffin sectioning are for reference only. Specific situations need to be assessed based on actual conditions.

Paraffin Section Common Problems: Causes and Solutions

Problem 1

Q: What should be done if there is overlapping and wrinkling in the sections?

A: This situation mainly falls into two categories. One is related to issues during sectioning, such as when the tissue is relatively hard, contains calcified tissues, or if the blade is not sharp enough during sectioning, leading to the formation of wrinkles. The other situation occurs when the paraffin block is not completely cooled, resulting in a soft texture, making it prone to wrinkling during sectioning. Another scenario is insufficient water temperature during dewaxing, preventing the complete unfolding of cut tissue sections, also causing the formation of wrinkles.

Solution: If the blade is not sharp enough, it needs to be replaced. For floating section, the water temperature should generally be 10°C lower than the melting point of paraffin. Adjusting the floating time allows the sections to fully expand. However, if the tissue originally contains some calcified parts, it may be challenging to avoid wrinkling.

 

Problem 2

Q: What should be done if there is scattered fragmentation in the sections?

A: This situation can be mainly attributed to two factors. One is when the water temperature is too high during floating, causing the pathological sections to spread out and further break apart. The other scenario is due to the high fat content in the sliced tissue itself, incomplete dehydration, especially in warm water, where fat is more likely to disperse.

Solution: It is necessary to promptly adjust the water temperature during floating, ensuring that the water temperature is suitable for the sections to spread out without overheating, which can lead to further fragmentation. Conversely, when the tissue has a high content of fat, the floating time should be shortened. If the sections are extremely prone to breaking, in order to preserve tissue integrity, it may be considered to skip the floating step.

Problem 3

Q: What should be done if there are obvious knife marks in the sections?

A: This situation is primarily caused by the presence of hard components in the tissue during sectioning, leading to the blade bouncing off and creating notches. Consequently, under microscopic observation, the tissue sections may exhibit defect-like patterns resembling elongated roadways.

Solution: In most cases, it is necessary to replace the blade to address this issue. Similarly, if there are calcified lesions or lymph node tissues in the tissue, these knife marks may be relatively unavoidable. If these knife marks do not affect staining observation and pathological diagnosis, they can be tolerated to a certain extent.

Problem 4

Q: What should be done if there are scratches or holes in the sections?

A: This situation is mainly caused by excessive speed during tissue sectioning, resulting in incomplete alignment and uneven cross-sections of the sections. Under microscopic observation, various-sized tissue voids may appear.

Solution: During tissue sectioning, it is crucial to control the cutting speed and ensure proper alignment. For rough trimming, it is recommended to set the thickness to 10μm, and during fine trimming, adjustments should be made to achieve a smooth paraffin block surface. This can help reduce the occurrence of voids, small white spots, and rough edges.

Problem 5

Q: What should be done if there is missing or omitted tissue in the sections?

A: This situation can mainly be divided into two categories. One is when the tissue is not completely leveled in the embedding box during the embedding process, resulting in the failure to cut the maximum cross-section during sectioning. The other situation occurs when the tissue is not evenly sampled during the sampling process, leading to missing or omitted sections in the slides, subsequently affecting the next steps of the experiment.

Solution: Each step of tissue processing should be standardized as much as possible. When sampling tissues, ensure a smooth section surface. During tissue embedding, every effort should be made to level each piece of tissue. Observing whether the paraffin block exposes the maximum cutting surface with each cut is also essential. If the desired result is not achieved, the tissue may need to be re-embedded.

Problem 6

Q: What should be done if there is obvious skipping of sections during sectioning?

A: This situation is mainly due to excess paraffin during the tissue embedding process, resulting in insufficient connection between the entire embedding box and the tissue. This leads to partial hollow spaces, causing variations in staining depth after staining.

Solution: In the event of such a situation, it is necessary to first check whether the slicing instrument is securely locked. Additionally, inspect the blade holder of the microtome to ensure it is not misaligned. If the issue is caused by inadequate or hollow paraffin blocks due to improper embedding, the tissue embedding process needs to be redone. When dealing with hard tissues, it may be necessary to reduce the cutting speed appropriately, allowing the paraffin block to warm up before resectioning.

 

Problem 7

Q: What should be done if there are cracks or fractures in the sections?

A: This situation often occurs in sample tissues such as organ membranes and certain lymphoid tissues. The causes of this issue are mainly related to the paraffin block’s slicing temperature being lower than the expected temperature. This leads to tissue fragility, and if the slicing speed is not well-controlled, slicing too quickly can also result in section fractures.

Solution: During sectioning, adjust the operating table temperature to 0°C and reduce the slicing speed. This can effectively reduce the occurrence of cracks or fractures. If conditions allow, using ice blocks at 0°C as a cooling medium for the paraffin block can also effectively minimize this issue.

Problem 8

Q: What should be done if there is whitening and difficulty in sectioning?

A: This situation occurs when the tissue is not adequately dehydrated, or the dehydration process is not effective. Consequently, paraffin does not fully penetrate the tissue, making sectioning difficult due to the lack of paraffin support.

Solution: In the presence of this situation, the freshness of the tissue should be considered first. It is recommended to start fixing the tissue within half an hour after removal from the body. The fixation time should be controlled between 8-16 hours, depending on the size of the specimen. It is advisable for the specimen volume not to exceed 1cm³. The specific dehydration conditions can be determined based on the experiment’s requirements.

Problem 9

Q: What should be done if there is water seepage during sectioning?

A: This situation is characterized by the expansion of the section and whitening of the surface when the paraffin block is left in place for an extended period, leading to the scrapping of the section. It may be due to incomplete dehydration of the tissue, allowing some water to seep back.

Solution: In most cases, this issue arises because the paraffin block has been left on the melted ice surface for too long. To prevent this situation, it is advisable to place the paraffin block on the ice surface for cooling again a few minutes before sectioning. This can significantly reduce the occurrence of water seepage during sectioning.

Problem 10

Q: What should be done if there is cracking of paraffin blocks during sectioning?

A: This situation primarily occurs because the temperature drops too rapidly during the cooling of the paraffin block, and the internal paraffin has not completely solidified. If the module is directly pried open under these conditions, it can lead to the paraffin block cracking.

Solution: To address this issue, it is necessary to analyze and resolve it from two perspectives. Firstly, check the quality of the paraffin. If there is a quality issue, replace the product. Secondly, during paraffin cooling, avoid rapid temperature drops. Allow the paraffin to cool naturally and completely solidify before further separation. This can effectively reduce the occurrence of paraffin block cracking.